Column-switching high-performance liquid chromatographic assay for the determination of quercetin in human urine with ultraviolet absorbance detection

Research output: Contribution to journalJournal articleResearchpeer-review

A high-performance liquid chromatographic method is described for the determination of quercetin in human urine using column-switching and ultraviolet (UV) absorbance detection. Urine samples were enzymatically hydrolysed and solid-phase extracted prior to injection onto the high- performance liquid chromatography (HPLC) system. Prior to elution of quercetin and the internal standard, fisetin, from the first column used for sample clean-up, the six-port valve was switched to the second column for analysis with UV detection. Detection of quercetin was accurate and reproducible, with a detection limit of 5 ng/ml. The method was applied to determine the urinary level of quercetin in 120 samples from an intervention study with fruit juice.

Original languageEnglish
JournalJournal of Chromatography B: Biomedical Sciences and Applications
Volume707
Issue number1-2
Pages (from-to)81-89
Number of pages9
ISSN0378-4347
DOIs
Publication statusPublished - 1998
Externally publishedYes

    Research areas

  • Quercetin

ID: 254770881